Help File
The program is to correct
the distortions caused by field inhomogeneity and eddy current induced field
because of strong diffusion weighting gradients in diffusion tensor imaging (DTI)
datasets acquired by echo planar imaging (EPI). All input volumes should be
Analyze format.
Calibration Tab
The calibration section
takes field maps of the base volume and at least four field maps in
non-coplanar diffusion weighing directions. The eddy current induced fields
will be saved to a file.
·
Raw Phase Map
File
1.
Select the radio
button to activate the group box. Click the GE button to select the raw P file
acquired by multi-echo phase volumes. The sequence is desribed
in NeuroImage xxxxxx.
2.
Click the
Calibration button to save the eddy current induced gradient file.
·
B0 and DW Field
Maps
1.
Select the radio
button to activate the group box. Click Browse button to select field variation
maps. The order of field map volume is b0, b1,b2,...,
bn.
2.
Click the Browse
button in "ECIF Estimation with Orthogonal Base" to select the
diffusion gradient directional scheme file. Each row in the scheme file
contains the normalized diffusion direction. The order should match the field
map volume order. An example of diffusion weighting directional scheme:
0.2160 -0.8986 0.3819
-0.4446 -0.4187 0.7918
-0.6352
-0.7701 -0.0595
0.5005 -0.1925
0.8440
-0.2212 0.5312
-0.8179
-0.8919 0.0791
-0.4452
… … …
3.
Click the
Calibrate button to save the eddy current induced field to disk.
Correct Tab
The correct section takes
the DTI file, the base field map and optional calibrated eddy current induced
field to correct the DTI file.
1.
Click the browse
button in DTI Input box to select the DTI dataset.
2.
Click the browse button
in B0 Map box to select the base field map.
3.
If no eddy
current induced field information, click the Correct
button to correct the distorted DTI dataset without eddy current induced
distortion correction.
4.
If eddy current
induced field exists, click the radio button.
5.
Select the
Calibration file and Diffusion Weighting Scheme. The scheme should have the
same diffusion weighing order of the DTI dataset.
6.
Input acquisition
time in milliseconds.
7.
Click Correct to
correct the distortion caused by field inhomogeneity and eddy current induced
field.